{"id":824,"date":"2026-03-30T06:33:39","date_gmt":"2026-03-30T06:33:39","guid":{"rendered":"https:\/\/www.syringetest.com\/?p=824"},"modified":"2026-03-30T06:33:40","modified_gmt":"2026-03-30T06:33:40","slug":"iso9626-annex-b","status":"publish","type":"post","link":"https:\/\/www.syringetest.com\/fr\/iso9626-annex-b","title":{"rendered":"Understanding ISO9626 Annex B: The Needle Tubing Stiffness Test Method"},"content":{"rendered":"<h2 class=\"wp-block-heading\">Introduction to ISO9626 Annex B<\/h2>\n\n\n\n<p>Le <strong>ISO9626 Annex B<\/strong> standard provides a comprehensive testing method for evaluating the stiffness of needle tubing used in medical devices. This test is essential for ensuring that the <strong>needle tubing<\/strong> maintains the necessary rigidity for safe and effective use in medical applications, such as syringes and injectables. By following this standard, manufacturers can guarantee that their products meet the required safety and performance criteria.<\/p>\n\n\n\n<p>Le <strong>essai de rigidit\u00e9 d'un tube d'aiguille<\/strong> outlined in Annex B of ISO9626 is a critical procedure for manufacturers of medical devices, particularly those dealing with syringes and other injectable products. This test helps in determining whether the needle tubing has the proper mechanical properties to function effectively without risk of failure during use.<\/p>\n\n\n\n<figure class=\"wp-block-image size-full is-resized\"><a href=\"https:\/\/www.syringetest.com\/fr\/standards\/iso-9626\/\" target=\"_blank\" rel=\" noreferrer noopener\"><img fetchpriority=\"high\" decoding=\"async\" width=\"600\" height=\"600\" src=\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp\" alt=\"ISO9626 Annex B Tubing Stiffness Test\" class=\"wp-image-826\" style=\"width:793px;height:auto\" srcset=\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp 600w, https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test-300x300.webp 300w, https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test-150x150.webp 150w, https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test-12x12.webp 12w\" sizes=\"(max-width: 600px) 100vw, 600px\" \/><\/a><\/figure>\n\n\n\n<h2 class=\"wp-block-heading\">What is ISO9626 Annex B?<\/h2>\n\n\n\n<p>ISO9626 is a standard that governs the <strong>manufacturing and testing of medical needle tubing<\/strong> used in the production of syringes and other medical devices. Annex B specifically focuses on the stiffness of the tubing, which is crucial for the performance of syringes and their ability to inject the contents reliably and safely.<\/p>\n\n\n\n<p>In Annex B, the testing procedure involves applying a force to the center of the tubing, which is supported at both ends, and measuring the deflection that occurs. This allows manufacturers to assess whether the tubing\u2019s stiffness meets the required specifications for use in syringes, needles, and similar medical devices.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Needle Tubing Stiffness Test: The Key Procedure<\/h2>\n\n\n\n<p>Le <strong>essai de rigidit\u00e9 d'un tube d'aiguille<\/strong> is performed using a specialized apparatus that applies a specific force to the tubing and measures its deflection. The test is carried out according to the guidelines provided in ISO9626 Annex B, which specifies the required parameters and equipment.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">3-Point Bend Test for Tubing Stiffness<\/h3>\n\n\n\n<p>Le <strong>3-point bend test<\/strong> is an essential part of the <strong>essai de rigidit\u00e9 d'un tube d'aiguille<\/strong>. In this procedure, the tubing is placed on a testing apparatus that applies a downward force to the center of the tubing. The <strong>la trav\u00e9e<\/strong> between the two supporting ends is adjusted according to the standard for the specific tubing size.<\/p>\n\n\n\n<p>The amount of deflection at the center of the tubing is measured, and this deflection value is used to assess the tubing&#8217;s stiffness. This test ensures that the <strong>needle tubing<\/strong> has the right balance of rigidity and flexibility, which is vital for the safety and functionality of syringes and other medical devices.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">The Apparatus for Needle Tubing Stiffness Test<\/h2>\n\n\n\n<p>Pour effectuer la <strong>essai de rigidit\u00e9 d'un tube d'aiguille<\/strong>, the apparatus must meet specific requirements outlined in ISO9626 Annex B:<\/p>\n\n\n\n<ul class=\"wp-block-list\">\n<li><strong>Plongeur<\/strong>: A blunt wedge-shaped plunger with a cylindrical radius of 1mm is used to apply a force on the tubing. The plunger should be able to apply a force of up to 60 N with an accuracy of \u00b10.1 N.<\/li>\n\n\n\n<li><strong>Deflection Measuring Equipment<\/strong>: The equipment used to measure the deflection should have a high level of precision, capable of measuring the deflection to the nearest 0.01 mm.<\/li>\n\n\n\n<li><strong>Support and Span<\/strong>: The tubing must be supported at both ends, and the span between the supports should be adjusted according to the specified dimensions for the tubing.<\/li>\n<\/ul>\n\n\n\n<h2 class=\"wp-block-heading\">Injectability Test and Its Importance<\/h2>\n\n\n\n<p>In addition to the <strong>essai de rigidit\u00e9 d'un tube d'aiguille<\/strong>, le <strong>test d'injectabilit\u00e9<\/strong> is another crucial evaluation for medical needles and syringes. This test assesses how easily the contents can be injected through the needle, ensuring that the needle does not cause excessive resistance or issues during the injection process.<\/p>\n\n\n\n<p>Le <strong>test d'injectabilit\u00e9<\/strong> often involves simulating real-world conditions by applying pressure to the syringe and measuring the force required to inject the contents. A needle that is too stiff may cause difficulty in injection, while one that is too flexible may lead to instability or potential failure.<\/p>\n\n\n\n<h2 class=\"wp-block-heading\">Role of Syringe Testing Equipment<\/h2>\n\n\n\n<p>For manufacturers adhering to <strong>ISO9626 Annex B<\/strong>, utilizing precise <strong>\u00e9quipement de test des seringues<\/strong> is essential. This equipment ensures that the <strong>essai de rigidit\u00e9 d'un tube d'aiguille<\/strong> et <strong>test d'injectabilit\u00e9<\/strong> are carried out accurately and consistently. High-quality testing instruments allow manufacturers to meet regulatory requirements and deliver safe, reliable products.<\/p>\n\n\n\n<p>For instance, the <strong>Tubing Stiffness Tester<\/strong> from <strong>Instruments cellulaires<\/strong> is designed to provide accurate stiffness measurements for needle tubing. This instrument is ideal for testing the <strong>stiffness of needle tubing<\/strong> used in syringes and other medical devices, ensuring compliance with <strong>ISO9626 Annex B<\/strong> and other relevant standards.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\"><\/h3>\n\n\n\n<p>Le <strong>ISO9626 Annex B<\/strong> standard plays a critical role in ensuring the safety and performance of needle tubing used in medical devices. The <strong>essai de rigidit\u00e9 d'un tube d'aiguille<\/strong> is vital for verifying that the tubing can withstand the forces encountered during use while maintaining the necessary flexibility for injection procedures.<\/p>\n\n\n\n<p>Manufacturers can ensure that their products meet the required performance standards by implementing the <strong>3-point bend test<\/strong> for tubing stiffness and incorporating reliable <strong>\u00e9quipement de test des seringues<\/strong> into their quality control processes. <strong>Instruments cellulaires<\/strong> offers advanced <strong>Tubing Stiffness Testers<\/strong> to assist manufacturers in meeting these stringent requirements and ensuring the quality of their medical products.<\/p>\n\n\n\n<p><\/p>","protected":false},"excerpt":{"rendered":"<p>Introduction to ISO9626 Annex B The ISO9626 Annex B standard provides a comprehensive testing method for evaluating the stiffness of needle tubing used in medical devices. This test is essential for ensuring that the needle tubing maintains the necessary rigidity for safe and effective use in medical applications, such as syringes and injectables. By following [&hellip;]<\/p>","protected":false},"author":4,"featured_media":0,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"footnotes":""},"categories":[1],"tags":[],"class_list":["post-824","post","type-post","status-publish","format-standard","hentry","category-uncategorized"],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v25.0 - https:\/\/yoast.com\/wordpress\/plugins\/seo\/ -->\n<title>ISO9626 Annex B Tubing Stiffness Test<\/title>\n<meta name=\"description\" content=\"Learn about ISO9626 Annex B testing standards for needle tubing stiffness. Discover the needle tubing stiffness test method\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.syringetest.com\/fr\/iso9626-annex-b\/\" \/>\n<meta property=\"og:locale\" content=\"fr_FR\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"ISO9626 Annex B Tubing Stiffness Test\" \/>\n<meta property=\"og:description\" content=\"Learn about ISO9626 Annex B testing standards for needle tubing stiffness. Discover the needle tubing stiffness test method\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.syringetest.com\/fr\/iso9626-annex-b\/\" \/>\n<meta property=\"og:site_name\" content=\"Syringe and Needle Testing\" \/>\n<meta property=\"article:published_time\" content=\"2026-03-30T06:33:39+00:00\" \/>\n<meta property=\"article:modified_time\" content=\"2026-03-30T06:33:40+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp\" \/>\n\t<meta property=\"og:image:width\" content=\"600\" \/>\n\t<meta property=\"og:image:height\" content=\"600\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/webp\" \/>\n<meta name=\"author\" content=\"Editor\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"\u00c9crit par\" \/>\n\t<meta name=\"twitter:data1\" content=\"Editor\" \/>\n\t<meta name=\"twitter:label2\" content=\"Dur\u00e9e de lecture estim\u00e9e\" \/>\n\t<meta name=\"twitter:data2\" content=\"4 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\/\/schema.org\",\"@graph\":[{\"@type\":\"Article\",\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b#article\",\"isPartOf\":{\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b\"},\"author\":{\"name\":\"Editor\",\"@id\":\"https:\/\/www.syringetest.com\/#\/schema\/person\/b09123bcbf540e0b55ec00028ad48511\"},\"headline\":\"Understanding ISO9626 Annex B: The Needle Tubing Stiffness Test Method\",\"datePublished\":\"2026-03-30T06:33:39+00:00\",\"dateModified\":\"2026-03-30T06:33:40+00:00\",\"mainEntityOfPage\":{\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b\"},\"wordCount\":822,\"commentCount\":0,\"publisher\":{\"@id\":\"https:\/\/www.syringetest.com\/#organization\"},\"image\":{\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b#primaryimage\"},\"thumbnailUrl\":\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp\",\"articleSection\":[\"Uncategorized\"],\"inLanguage\":\"fr-FR\",\"potentialAction\":[{\"@type\":\"CommentAction\",\"name\":\"Comment\",\"target\":[\"https:\/\/www.syringetest.com\/iso9626-annex-b#respond\"]}]},{\"@type\":\"WebPage\",\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b\",\"url\":\"https:\/\/www.syringetest.com\/iso9626-annex-b\",\"name\":\"ISO9626 Annex B Tubing Stiffness Test\",\"isPartOf\":{\"@id\":\"https:\/\/www.syringetest.com\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b#primaryimage\"},\"image\":{\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b#primaryimage\"},\"thumbnailUrl\":\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp\",\"datePublished\":\"2026-03-30T06:33:39+00:00\",\"dateModified\":\"2026-03-30T06:33:40+00:00\",\"description\":\"Learn about ISO9626 Annex B testing standards for needle tubing stiffness. Discover the needle tubing stiffness test method\",\"breadcrumb\":{\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b#breadcrumb\"},\"inLanguage\":\"fr-FR\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\/\/www.syringetest.com\/iso9626-annex-b\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"fr-FR\",\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b#primaryimage\",\"url\":\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp\",\"contentUrl\":\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\/\/www.syringetest.com\/iso9626-annex-b#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\/\/www.syringetest.com\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Understanding ISO9626 Annex B: The Needle Tubing Stiffness Test Method\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\/\/www.syringetest.com\/#website\",\"url\":\"https:\/\/www.syringetest.com\/\",\"name\":\"Syringe and Needle Testing\",\"description\":\"Syringe and Needle Test Solutions\",\"publisher\":{\"@id\":\"https:\/\/www.syringetest.com\/#organization\"},\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\/\/www.syringetest.com\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"fr-FR\"},{\"@type\":\"Organization\",\"@id\":\"https:\/\/www.syringetest.com\/#organization\",\"name\":\"Syringe and Needle Testing\",\"url\":\"https:\/\/www.syringetest.com\/\",\"logo\":{\"@type\":\"ImageObject\",\"inLanguage\":\"fr-FR\",\"@id\":\"https:\/\/www.syringetest.com\/#\/schema\/logo\/image\/\",\"url\":\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2025\/04\/syringetest.webp\",\"contentUrl\":\"https:\/\/www.syringetest.com\/wp-content\/uploads\/2025\/04\/syringetest.webp\",\"width\":512,\"height\":512,\"caption\":\"Syringe and Needle Testing\"},\"image\":{\"@id\":\"https:\/\/www.syringetest.com\/#\/schema\/logo\/image\/\"}},{\"@type\":\"Person\",\"@id\":\"https:\/\/www.syringetest.com\/#\/schema\/person\/b09123bcbf540e0b55ec00028ad48511\",\"name\":\"Editor\",\"image\":{\"@type\":\"ImageObject\",\"inLanguage\":\"fr-FR\",\"@id\":\"https:\/\/www.syringetest.com\/#\/schema\/person\/image\/\",\"url\":\"https:\/\/secure.gravatar.com\/avatar\/95df1d699cb620958de379e8fcffd6ab3f1693eafdfbb86ed0594de528b77bb0?s=96&d=mm&r=g\",\"contentUrl\":\"https:\/\/secure.gravatar.com\/avatar\/95df1d699cb620958de379e8fcffd6ab3f1693eafdfbb86ed0594de528b77bb0?s=96&d=mm&r=g\",\"caption\":\"Editor\"},\"url\":\"https:\/\/www.syringetest.com\/fr\/author\/editor\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"ISO9626 Annex B Tubing Stiffness Test","description":"Learn about ISO9626 Annex B testing standards for needle tubing stiffness. Discover the needle tubing stiffness test method","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.syringetest.com\/fr\/iso9626-annex-b\/","og_locale":"fr_FR","og_type":"article","og_title":"ISO9626 Annex B Tubing Stiffness Test","og_description":"Learn about ISO9626 Annex B testing standards for needle tubing stiffness. Discover the needle tubing stiffness test method","og_url":"https:\/\/www.syringetest.com\/fr\/iso9626-annex-b\/","og_site_name":"Syringe and Needle Testing","article_published_time":"2026-03-30T06:33:39+00:00","article_modified_time":"2026-03-30T06:33:40+00:00","og_image":[{"width":600,"height":600,"url":"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp","type":"image\/webp"}],"author":"Editor","twitter_card":"summary_large_image","twitter_misc":{"\u00c9crit par":"Editor","Dur\u00e9e de lecture estim\u00e9e":"4 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"Article","@id":"https:\/\/www.syringetest.com\/iso9626-annex-b#article","isPartOf":{"@id":"https:\/\/www.syringetest.com\/iso9626-annex-b"},"author":{"name":"Editor","@id":"https:\/\/www.syringetest.com\/#\/schema\/person\/b09123bcbf540e0b55ec00028ad48511"},"headline":"Understanding ISO9626 Annex B: The Needle Tubing Stiffness Test Method","datePublished":"2026-03-30T06:33:39+00:00","dateModified":"2026-03-30T06:33:40+00:00","mainEntityOfPage":{"@id":"https:\/\/www.syringetest.com\/iso9626-annex-b"},"wordCount":822,"commentCount":0,"publisher":{"@id":"https:\/\/www.syringetest.com\/#organization"},"image":{"@id":"https:\/\/www.syringetest.com\/iso9626-annex-b#primaryimage"},"thumbnailUrl":"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp","articleSection":["Uncategorized"],"inLanguage":"fr-FR","potentialAction":[{"@type":"CommentAction","name":"Comment","target":["https:\/\/www.syringetest.com\/iso9626-annex-b#respond"]}]},{"@type":"WebPage","@id":"https:\/\/www.syringetest.com\/iso9626-annex-b","url":"https:\/\/www.syringetest.com\/iso9626-annex-b","name":"ISO9626 Annex B Tubing Stiffness Test","isPartOf":{"@id":"https:\/\/www.syringetest.com\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.syringetest.com\/iso9626-annex-b#primaryimage"},"image":{"@id":"https:\/\/www.syringetest.com\/iso9626-annex-b#primaryimage"},"thumbnailUrl":"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp","datePublished":"2026-03-30T06:33:39+00:00","dateModified":"2026-03-30T06:33:40+00:00","description":"Learn about ISO9626 Annex B testing standards for needle tubing stiffness. Discover the needle tubing stiffness test method","breadcrumb":{"@id":"https:\/\/www.syringetest.com\/iso9626-annex-b#breadcrumb"},"inLanguage":"fr-FR","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.syringetest.com\/iso9626-annex-b"]}]},{"@type":"ImageObject","inLanguage":"fr-FR","@id":"https:\/\/www.syringetest.com\/iso9626-annex-b#primaryimage","url":"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp","contentUrl":"https:\/\/www.syringetest.com\/wp-content\/uploads\/2026\/03\/ISO9626-Annex-B-Tubing-Stiffness-Test.webp"},{"@type":"BreadcrumbList","@id":"https:\/\/www.syringetest.com\/iso9626-annex-b#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/www.syringetest.com\/"},{"@type":"ListItem","position":2,"name":"Understanding ISO9626 Annex B: The Needle Tubing Stiffness Test Method"}]},{"@type":"WebSite","@id":"https:\/\/www.syringetest.com\/#website","url":"https:\/\/www.syringetest.com\/","name":"Syringe and Needle Testing","description":"Syringe and Needle Test Solutions","publisher":{"@id":"https:\/\/www.syringetest.com\/#organization"},"potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/www.syringetest.com\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"fr-FR"},{"@type":"Organization","@id":"https:\/\/www.syringetest.com\/#organization","name":"Syringe and Needle Testing","url":"https:\/\/www.syringetest.com\/","logo":{"@type":"ImageObject","inLanguage":"fr-FR","@id":"https:\/\/www.syringetest.com\/#\/schema\/logo\/image\/","url":"https:\/\/www.syringetest.com\/wp-content\/uploads\/2025\/04\/syringetest.webp","contentUrl":"https:\/\/www.syringetest.com\/wp-content\/uploads\/2025\/04\/syringetest.webp","width":512,"height":512,"caption":"Syringe and Needle Testing"},"image":{"@id":"https:\/\/www.syringetest.com\/#\/schema\/logo\/image\/"}},{"@type":"Person","@id":"https:\/\/www.syringetest.com\/#\/schema\/person\/b09123bcbf540e0b55ec00028ad48511","name":"Editor","image":{"@type":"ImageObject","inLanguage":"fr-FR","@id":"https:\/\/www.syringetest.com\/#\/schema\/person\/image\/","url":"https:\/\/secure.gravatar.com\/avatar\/95df1d699cb620958de379e8fcffd6ab3f1693eafdfbb86ed0594de528b77bb0?s=96&d=mm&r=g","contentUrl":"https:\/\/secure.gravatar.com\/avatar\/95df1d699cb620958de379e8fcffd6ab3f1693eafdfbb86ed0594de528b77bb0?s=96&d=mm&r=g","caption":"Editor"},"url":"https:\/\/www.syringetest.com\/fr\/author\/editor"}]}},"_links":{"self":[{"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/posts\/824","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/users\/4"}],"replies":[{"embeddable":true,"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/comments?post=824"}],"version-history":[{"count":2,"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/posts\/824\/revisions"}],"predecessor-version":[{"id":827,"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/posts\/824\/revisions\/827"}],"wp:attachment":[{"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/media?parent=824"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/categories?post=824"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/www.syringetest.com\/fr\/wp-json\/wp\/v2\/tags?post=824"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}